Model | BIOBASE2000 |
Sample Unit | Sample Racks | Standard 192 sample positions(expandable, up to 552) |
Pipetting | 10~1000μl |
Dispensing System | 2 aspirating and dispensing probe( X-Y-Z movement) |
Reagent Unit
| Reagent Position | 23 reagent racks, editable |
Pipetting Range | 10~1000μl, 1μl stepping |
Pipetting Precision | (100μl)CV≤0.5% |
Washing Unit
| Dispensing Time | 4 minutes to whole 96 to microplates |
Washing Probes | Two line 8-nozzle manifolds |
Wash Containers | Cleaning fluid(15L), waste water(25L), Buffer 1(5L), Buffer 2(2L), Buffer 3(2L), Buffer 4(2L), with liquid level-sensing, liquid shortage and full alarm |
Waste Containers | 25L with waste full sensor |
Washing Residual | <2μl |
Reading Unit
| Reading Channels | 8 independent photometric channels, mono anddichromatic reading |
Absorbance | 0.0~3.0OD |
Spectral Range | 400~700nm |
Optical Filters | 2 standard filters(450, 630nm); 2 more on request (405nm/450nm/492nm/630nm) |
Reading Accuracy | ±1% from 0~1.5OD, ±2% from 1.5~3.0OD |
Incubation Unit | Incubators | 4 independent units |
Temperature Range | RT to 45℃ |
Temperature Accuracy | ±0.5℃ |
Data Management | Software System | Above Windows 7 system |
LIS System | Bi-direction, support HL7 protocol |
Results | Absorbance and results reviewable by software |
Printer | External optional |
Working Conditions | Power Supply | AC220V±10%, 50/60Hz; 110V±10%, 60Hz, 400W |
Temperature | 10~30℃ |
Humidity | 30%~80% |
Size&Weight | External Size (W*D*H) | 1235*705*875mm |
Net Weight | 146kg |
Package Size (W*D*H) | 1380*860*1435mm |
Gross Weight | 220kg |